Calibrate Sensor Data
Apply a confirmed temperature or humidity offset to a Sensor in the MOCREO Sensor App.
Can I calibrate my MOCREO sensor?
-
However, we generally do not recommend that you calibrate, because our MOCREO Sensor is strictly inspected and tested before leaving the factory, and the measurement error of the Sensor is usually within ±0.9°F (±0.5°C). At the same time, we attach the sensor chip manual for your reference.
If you have any other questions, please contact us(https://mocreo.com/service)
-
Yes.
-
First, you need an accurate temperature and/or humidity reference. Each value is calibrated separately, and you can choose to calibrate one, both, or none as needed. The accuracy of a non-calibrated sensor is set out in the complete specifications.
-
The available compensation range is -5.0 to +5.0°C (-9.0 to +9.0°F) for temperature and -5.0 to +5.0% RH for humidity. The step is 0.1°C or 0.1% RH; in Fahrenheit, each step is 0.18°F. See Temperature and Humidity Compensation for Sensor App alert and historical-data behavior.
-
Calibration should only correct a confirmed fixed variation. If you find a larger or unstable difference, please contact us, as this may indicate a problem with the calibration process or the Sensor.
-
The following situations may cause larger temperature/humidity differences:
- Sensor hardware damage (verification method: when you have multiple sensors, you can place multiple sensors in the same environment and measure differences of ±1.8°F (±1°C) is normal)