Skip to content

Sensor Calibration and Certification

Do MOCREO Sensors Require Re-calibration After Purchase?

No, MOCREO sensors do not require re-calibration after purchase. All MOCREO sensors undergo rigorous accuracy and consistency testing before shipment, so you can use them directly upon receipt without additional calibration.

Verification Testing (Optional)

If you want to verify your sensor's accuracy, you can perform a comparison test:

  1. Deploy two identical sensors in the same location
  2. For sensors with external probes: Position the probes side by side
  3. Wait 1-2 days and observe the monitoring data
  4. Compare the data fluctuations between the two sensors
  5. Analyze any differences to determine if issues are environmental or sensor-related

Consistent readings indicate accurate sensors with environment-related variations. Significant differences may suggest sensor malfunction requiring support contact.

Does MOCREO Provide Calibration Certificates?

MOCREO does not issue individual sensor calibration certificates because our customer base covers too many use cases with varying calibration requirements. However, you can have your sensors calibrated by third-party calibration services.

Third-Party Calibration Services

You can send your MOCREO sensors to certified calibration laboratories that provide NIST-traceable certification. These services offer both commercial and certified calibration with full documentation.

Recommended Calibration Providers:

Important Notes: - Pricing: Contact providers directly for current pricing (this is not an inexpensive process) - Certification Types: Both commercial and certified calibration are available - NIST Traceability: All services provide NIST-traceable certification - Documentation: Calibration certificates include original data and test results

MOCREO Calibration Statement

If you need official documentation, MOCREO can provide a MOCREO Calibration Statement confirming that specified sensor models undergo rigorous calibration before shipment.

Need Additional Information?

For questions about sensor calibration or certification requirements, please contact our support team with: - Your specific calibration needs - Intended use case or industry requirements - Sensor model and quantity - Any regulatory compliance requirements