Calibrate Sensor Data

Path: Home → Device card → Settings → Data Calibration

Calibrate a Reading

Path: Home → Device card → Settings → Data Calibration

  1. Place an accurate reference instrument beside the sensor and allow both readings to stabilize.
  2. Adjust the available temperature or humidity offset.
  3. Tap Save.

The available offset range is -5.0 to +5.0°C (-9.0 to +9.0°F) for temperature and -5.0 to +5.0% RH for humidity, with a step of 0.1 in the current unit. See Temperature and Humidity Compensation for multi-probe field mapping and historical-data behavior.

Calibration is intended for a confirmed fixed difference. A large or unstable difference should be investigated instead of hidden with an offset.